OBJ_BUCH-0000000505-010.book Seite 10 Freitag, 8. Juni 2018 12:28 12
Purpose
en
10
WARNING!
Replacing fuse
Replace the fuse only with a fuse of the same type and rating (see Chapter 11,
Technical Data).
WARNING!
Equipment should be positioned such that power cord can be easily disconnect-
ed.
WARNING!
Electromagnetic emissions may increase and rise above the permissible limits if
other equipment (e.g. MyoSure® Control Unit) is stacked onto or placed directly
next to the Aquilex Fluid Control System. The user is responsible for monitoring
the devices to make sure they function properly.
WARNING!
Electrical Interference (see Chapter 12, Guidelines and Manufacturer's State-
ment - Electromagnetic Compatibility): Electrical interference with other devices
or instruments was considered when developing this system and none was de-
tected during testing. However, if you still detect or suspect such interference,
please follow these suggestions:
• Move the Aquilex System, the other device, or both devices to a different loca-
tion
• Increase distance between devices used
• Consult an electro-medical expert
WARNING!
The Aquilex Fluid Control System should not be used directly next to other devic-
es as this could result in malfunctions. The Aquilex Fluid Control System was test-
ed for compliance with IEC 60601-1-2 as a stand alone system. Therefore, do not
stack other devices (e.g. MyoSure® Control Unit) on the system or the Irrigation
Pump Unit. In particular, do not place any other device than the AQL-100PBS on
the trays of the AQL-100CBS. If usage in the manner described above is neverthe-
less required, this system and the other devices should be monitored to make
sure they function properly.
WARNING!
If the Aquilex Fluid Control System is configured as part of a ME SYSTEM, the en-
tire ME SYSTEM should be tested for compliance with IEC 60601-1-1, and any
equipment used with the Aquilex Fluid Control System should be Type BF.
WARNING!
If the leakage current of the configured ME SYSTEM exceeds the limits of
IEC 60601-1-1, install an appropriately rated UL 2601-1/IEC 60601-1 approved
isolation transformer and retest the system.