System Operation
en
Fig. 5-7
Fluid bag suspension
(1)
Bag scale
Fluid bag on bag hook
(2)
Bag deflector
(3)
22
3. If a tube set is in the inflow tube holder when the pump is switched on, the
pump display (Fig. 5-1, Front of Irrigation Pump Unit (1)) shows the message
Remove Tube Set. The device self-test resumes once the tube set is removed
from the roller wheel.
If the device self-test is unsuccessful, the corresponding error messages are
displayed (see Chapter 10, Error and Warning Messages).
The system has successfully completed the device self-test when a single audible
beep is heard. The message System OK is displayed for 5 secs followed by the
message Insert Tube Set.
WARNING!
Do not use this system if a defect is suspected or detected during the function
check. This also applies to obvious defects, especially defects and damage to the
power plug and power cord.
5.5
Hanging the Fluid Bags
WARNING!
When performing monopolar hysteroscopic electrosurgery, the distension medi-
um must be electrically non-conductive. Examples include glycine, sorbitol and
mannitol. Isotonic saline irrigation fluids may only be used when performing bi-
polar electrosurgical resective procedures.
Hang one or two fluid bags filled with distension fluid appropriate for procedure.
WARNING!
The system is only intended for use with flexible fluid containers. Do not use
glass containers as they might break. With rigid containers, fluid cannot flow
quickly enough due to the vacuum being generated inside of the containers. Risk
of implosion with rigid containers.