Functional Check

WARNING!
Functional test
The functional test must be performed prior to each device use.
1. Perform a visual check of the devices. Do not use the system in case of obvi-
ous damage.
2. Check the rollers of the roller wheel to make sure they move easily and
smoothly.
3. Turn on the device, check whether power switch and indicators and displays
light up accordingly.
4. The device self-test must run successfully; there are no displayed error mes-
sages (5.4, Turning On the Aquilex System).
5. The bags with the irrigation fluid must hang freely and may not touch the
scale.
6. Check that all tube connections (vacuum/inflow/outflow) were made cor-
rectly and are intact.
7. Check to make sure all tube connections are free of mechanical stresses and
are routed without snagging. The tube connections may not touch the scale.
Non-observance can lead to a distortion of the deficit calculation.
8. The automatic lumen calibration has been carried out successfully; there are
no displayed error messages (5.12, Using the Pump during Surgery).
9. Check for leaking irrigation fluid in the area of the pressure chamber.
Functional Check