OBJ_BUCH-0000000505-010.book Seite 47 Freitag, 8. Juni 2018 12:28 12
12.3
Guidelines and Manufacturer's Statement/Electromagnetic In-
terference Immunity
TheAquilex Fluid Control System is intended for use in an electromagnetic envi-
ronment as described below. The user/operator of the Aquilex Fluid Control Sys-
tem must make sure the device is operated within such an environment.
Electromagnetic
IEC 60601-1-2 test level
interference
immunity tests
Discharge of static
± 8 kV contact ± 15 kV air
electricity (ESD)
according to
IEC 61000-4-2
Electrical fast
± 2 kV for power lines
transients/bursts
± 1 kV for input and out-
according to
put lines
IEC 61000-4-4
Modulation 100 KHz
Voltage surges
± 1 kV line(s) to line(s)
according to
± 2 kV line(s) to earth
IEC 61000-4-5
Blackouts, brown-
< 5 % U
outs, and fluctua-
the U
tions of the power
< 40% U
supply according
U
to IEC 61000-4-11
T
< 70% U
U
T
< 5% U
U
T
Supply frequency
30 A/m
magnetic field
(50/60 Hz) accord-
ing to
IEC 61000 4-8
*Note: U
is the mains alternating voltage before applying the test levels.
T
Guidelines and Manufacturer's Statement - Electromagnetic Compatibility
Compliance Levels
± 8 kV contact
± 15 kV air
± 2 kV for power lines
± 1 kV for input and
output lines
Modulation 100 KHz
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
* (> 95 % dip in
< 5 % U
T
T
) for ½ cycle
in the UT) for 1/2
T
cycle
(> 60% dip in the
40 % U
(60 % dip in
T
T
) for 5 cycles
the U
) for 5 cycles 70
T
% U
(30 % dip in the
(> 30% dip in the
T
T
U
) for 25 cycles < 5 %
) for 25 cycles
T
U
(> 95 % dip in the
T
(> 95% dip in the
U
) for 5 seconds 0 %
T
T
) for 5 cycles
U
; 1/2 cycle at 0°,
T
45°, 90°, 135°, 180°,
225°,270°, and 315°
0% U
; 1 cycle and
T
70% U
; 25/30 cycles
T
Single phase: at 0°
0% U
; 250/300
T
cycles
30 A/m
Electromagnetic environ-
ment/guidelines
Floors should be made
from wood or concrete or
covered with ceramic
tiles. If the floor covering
consists of synthetic
material, the relative
humidity should be at
least 30 %.
The quality of the supply
voltage should be the
same as the voltage of a
typical business or hospi-
tal environment.
The quality of the supply
voltage should be the
same as the voltage of a
typical business or hospi-
tal environment.
* (> 95 % dip
The quality of the supply
voltage should be the
same as the voltage of a
typical business or hospi-
tal environment. If the
user/operator of the sys-
tem requires the contin-
uation of functionality
after power interrup-
tions/disruptions, it is
recommended to supply
the device with power
from an uninterruptible
power supply.
Magnetic fields of the
mains power frequency
should comply with the
typical values of business
and hospital environ-
ments.
en
47