Herunterladen Inhalt Inhalt Diese Seite drucken

Determine The Software Version - Hologic Aquilex Fluid Control System Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für Aquilex Fluid Control System:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 66
OBJ_BUCH-0000000505-010.book Seite 40 Freitag, 8. Juni 2018 12:28 12
Annual Inspection
en
Fig. 9-4
Vacuum pump exhaust ports
40
TM
TYPE
Aquilex Fluid Control System
L
SN / REF
Ser. Nr. WOM
100-240 V / 50-60 Hz
O
YYYY-MM
W
9.3

Determine the Software Version

The software version of the pump may be required for further analysis.
Follow these steps to determine the software version:
1. Turn device on and wait for the device self-test.
2. Press the Pause/Resume button for at least 2 seconds.
3. The software version of the pump is depicted in the "Deficit Limit" display and
consists of a 5-digit numerical sequence.
Enter the determined software version into the test log in Section 16.1.