System Operation
en
Instrument change during surgery

Total inflow volume displayed

Turning Off
30
WARNING!
A container change during surgery is only allowed, if the container holds at least
0,5 liters of fluid. Otherwise, the deficit value may be falsified. In this case, the
manufacturer recommends manual deficit calculation.
WARNING!
Touching the containers and their holders as well as vibrations of the balancing
system should be avoided during surgery to prevent false detection of the con-
tainer change and not negatively affect the accuracy of the deficit calculation.
WARNING!
Containers should be changed quickly to avoid affecting the accuracy of the def-
icit calculation.
CAUTION!
To avoid affecting the accuracy of the deficit calculation ensure that the first step
of the container change is to disconnect tubing from the full containers. Remove
full containers from the scale immediately after that.
5.15

Changing Instrument during Surgery

Pause the pump by pressing Pause/Resume button.
Press the Prime button for 2 seconds.
Change the instrument.
Fully open hysteroscope inflow stopcock.
Keep the hysteroscope at the height of the patient and above the drape to
ensure the fluid can be collected. Do not insert the hysteroscope into the
uterus at this time.
Press the Prime (17) button (Fig. 5-1).
Pump is running to carry out the automatic lumen calibration. The pump dis-
play depicts Calibration Running.
Three beeps are heard once the automatic lumen calibration is finished.
The pump display depicts Prime Successful Close Stopcock for 5 seconds, fol-
lowed by System Operating.
Close the stopcock for the hysteroscope inflow to stop the inflow.
5.16
Total Inflow Volume Displayed
If a manual check of the fluid deficit is desired, the total fluid volume supplied
from the fluid bags can be obtained by simultaneously pressing and holding both
the up and down arrows ((10) and (11) Fig. 5-1 ). The number in the Deficit Dis-
play is the total inflow fluid volume in ml. Once one or both of these buttons is
released, the Deficit Display will return to the Fluid Deficit value.
5.17

Turning System Off

Press the ON/OFF switch to turn pump off. The displays and indicators are no lon-
ger illuminated.
WARNING!
Pressing the ON/OFF switch does not disconnect the system from the wall power
outlet. This requires pulling the power cord located in the rear of the system.