• The device usage time function on
the aView
monitor is not available
TM
for VivaSight-SL. For this device, the
device usage time will always display
"00:00:00". This will not affect the
functioning of the system.
EXTUBATION
• Deflate the cuff completely, as
per protocol.
• Extubation is performed slowly, while
examining the airway on the aView
monitor one last time (document or
record, as necessary).
DISCONNECTING VIVASIGHT-SL
1. Disconnect the single use adapter
cable from the aView
2. Turn off the aView
pressing the power button for at least
two seconds.
3. Do not use the same VivaSight-SL for
more than 14 days and do not turn
the VivaSight-SL camera ON for more
than 8 hours in total. During long term
use of the VivaSight-SL, use the depth
marks on the VivaSight-SL to monitor
the tube position.
DISPOSAL
Discard the single use VivaSight-SL tube,
stylet and adapter cable in a suitable
biohazard receptacle in accordance with
local regulations.
For disposal instructions for the aView
monitor, refer to aView
instructions for use.
MANUFACTURED BY:
ETView Ltd.
Catom 2 Street
Misgav Business Park
M.P. Misgav 2017900, Israel
www.etview.com
LB10016 REV03
™
™
monitor.
™
monitor by
100%
™
™
monitor
Manufacture
Date
Consult instructions for use
Do not use if package is
damaged
Do not reuse
Sterilized Using
Ethylene Oxide
Non-sterile (single use
adapter cable only)
US Federal law restricts
this device to sale by or on
the order of a physician
Defibrillation-proof type BF
applied part
MR
Conditional
This product is not made
REF LOT
with phthalates
Temperature for
transportation and storage
Humidity for transportation
and storage
109kPa
Pressure for transportation
and storage
Authorized Representative
EC REP
in the European
Community
Waste Bin Symbol
REF LOT
This product is not made
with natural rubber latex
REF
Reference Number
Lot Number, Batch Code
Expiry
Date
Caution
STERILIZE
Manufacturer
STERILIZE
9