ENGLISH
Guidance and manufacturer's declaration
The system is intended for use in the electromagnetic environment specified
below. The customer or the user of the system should assure that it is used in
such an environment.
Immunity test IEC 60601-1-2
test level
Conducted
3 V RMS
Radio
0,15 MHz –
Frequency
80 MHz
IEC 61000-4-6
6 V RMS
in ISM bands
80% AM at 1 kHz
Radiated
3 V/m 80-2700
Radio
MHz
Frequency
80% AM at 1 kHz
IEC 61000-4-3
Additional
proximity field
test frequencies
and levels
according to
60601-1-2 table 9
NOTE 1 At 80 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
12
– electromagnetic immunity
Compliance level Electromagnetic
3 V RMS
0,15 MHz –
80 MHz
6 V RMS
in ISM bands
80% AM at 1 kHz
3 V/m 80-2700
MHz
80% AM at 1 kHz
Additional
proximity field
test frequencies
and levels
according to
60601-1-2 table 9
environment - guidance
Portable and mobile RF
communications equipment
should be used no closer
to any part of the system,
including its cables, than the
recommended separation
distance calculated from
the equation applicable
to the frequency of the
transmitter.
Recommended separation
distance
d = 1.17√ P
d = 1.17√ P 80 MHz to
800 MHz
d = 2.33√ P 800 MHz to
2.7 GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, a) should be less
than the compliance level in
each frequency range b).
Interference may occur in
the vicinity of equipment
marked with the
following symbol.