The Altrix system is intended for use in the electromagnetic environment specified below. The customer or the user
of Altrix should assure that it is used in such an environment.
Emissions test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations
Flicker Emissions
IEC 61000-3-3
Note: The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR
11 class A). If it is used in a residential environment, for which CISPR 11 class B is normally required, this equipment
might not offer adequate protection to radio frequency communication services. The user might need to take mitigation
measures, such as relocating or reorienting the equipment.
Recommended separations distances between portable and mobile RF communications equipment and the
The Altrix system is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of Altrix can help prevent electromagnetic interferences by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the Altrix system
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
1-58
EMC Information
Guidance and manufacturer's declaration - electromagnetic emissions
Compliance
220-240V/50Hz
220V/60Hz
Does not apply to 100V 50/
60Hz or 120V/60Hz
Complies
220-240V/50Hz only
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
D=(1.2) (√P)
0.12
0.38
1.2
3.8
8001-109-005 REV B
Electromagnetic environment
The Altrix system uses RF energy only for its
internal function. Therefore, its RF emissions
Group 1
are very low and are not likely to cause any
interference in nearby electronic equipment.
Class A
Class A
The Altrix system is suitable for use in all
establishments other than domestic and
those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Altrix system
m
80 MHz to 800 MHz
D=(0.35) (√P)
0.035
0.11
0.35
1.1
800 MHz to 2.7 GHz
D=(0.70) (√P)
0.07
0.22
0.7
2.2
www.stryker.com