Problem
Patient temperature
Patient temperature output
(PTO)
www.stryker.com
Troubleshooting
Possible cause
Out of range
External device output
displays high value >45° C
when input is out of range
(As a result of one of the
following: patient probe
disconnected, controller in
standby / sleep mode,
patient temperature outside
the range of 25° - 45°C).
8001-109-005 REV B
Action
Check the placement of
the probe
To resume calibration:
Disconnect the
external device from
the reusable adapter
cable
Tap the Help button to
display the alarm
screens
Find the "Temperature
Output Alarm" screen
Tap Confirm to restart
calibration
Wait until the "Monitor"
is solid
Connect the external
device to the reusable
adapter cable
Remove from use
Not applicable
Not applicable
1-51