Problem
Patient temperature
Patient temperature output
(PTO)
www.stryker.com
Troubleshooting
Possible cause
Out of range
External device output
displays high value >45° C
when input is out of range
(As a result of one of the
following: patient probe
disconnected, controller in
standby / sleep mode,
patient temperature outside
the range of 25° - 45°C).
8001-109-005 REV B
Action
Check the placement of
the probe
To resume calibration:
•
Disconnect the
external device from
the reusable adapter
cable
•
Tap the Help button to
display the alarm
screens
•
Find the "Temperature
Output Alarm" screen
•
Tap Confirm to restart
calibration
•
Wait until the "Monitor"
is solid
•
Connect the external
device to the reusable
adapter cable
Remove from use
Not applicable
Not applicable