Check water flow alarm (Continued)
EN
The Check Water Flow Alarm for the given outlet port stops, if the operator confirms removal.
The audio paused timeout is 10 minutes for the check water flow alarm.

No water flow alarm

This alarm notifies the operator of water flow that is not optimal.
Alarm generation:
When all 3 ports have a flow lower than 0.6 l/min for a minimum of 20 seconds, and the pump current is greater
than or equal to 1.0A and the pump is at full speed.
Notes
The pump continues to operate when the No Flow Alarm is active.
The audio paused timeout is 5 minutes for the no water flow alarm.

No water alarm

This alarm notifies the operator if there is no or low water in the system and the pump has shut down.
Alarm generation:
Pump is active in the Manual or Automatic mode. The pump current is lower than 1.0 A and 1 of the ports has a low
flow for more than 60 seconds.
Pump is active in the Manual or Automatic mode. The pump current is lower than 1.0 A and all 3 port flow states are
no flow for more than 60 seconds.
The pump current is less than 0.5 A for 20 seconds.
Note: The audio paused timeout is 5 minutes for the no water alarm.

Power backup level alarm

This alarm notifies to the operator an indicator of the status of the Backup Power Level.
Notes
The indicator will remain active until a qualified technician replaces the battery.
There is no reduction in the usability of the product. The product remains functional and a visual alarm is displays.
The product will disable the Backup Power Level Alarm in Sleep Mode. Otherwise, the alarm is enabled.
Alarm generation:
The product will display Backup Power Level Alarm when the battery level backup power is less than 100 minutes of
alarms. Once activated, the Backup Power Level Alarm will remain active until you power off the product.
Note: The audio paused timeout is 10 minutes for the power backup level alarm.

Therapy paused time out alarm

This alarm converts a therapy pause into an alarm if the duration of the pause is too long.
1-56
Alarm conditions