Disinfecting external surfaces (Continued)
•
Hose and power cord management straps
•
Storage compartment door
•
Inside storage compartment
•
Graphical user interface display
•
Handle
14. Follow specified contact time in accordance with the disinfectants manufacturer's instructions for use.
15. To rinse, wipe the hoses and patient temperature probe cables, with a soft, lint free cloth moistened with sterile
distilled water.
16. To rinse, wipe the controller, reservoir and reservoir lid surfaces, and system components with a lint free cloth
moistened with sterile distilled water.
17. To dry, wipe the controller, reservoir, reservoir lid surfaces, and system components with a clean, dry cloth to
remove any excess liquid.
18. Replace the reservoir.
19. Allow the external surfaces of the controller and components to dry thoroughly.
20. Store the power cord, cables, and hoses.
Disinfect the internal water circuit and hoses every 14 days
Use the BruClean TbC disinfectant tablets manufactured by Brulin (EPA registration number 71847-2-106) or
equivalent, before first use, at least every 14 days, and before storage. BruClean TbC has been validated for internal
water circuit disinfection. Make sure that you follow the disinfectant manufacturer's guidelines to avoid the risk of injury.
Failure to follow the disinfectant's instructions may void your warranty.
CAUTION
•
Always use sterile distilled water or distilled water that has been passed through a filter less than or equal to 0.22
microns with this product.
•
Do not disinfect the internal water system with a thermal transfer device attached as this may cause a leak.
•
Do not use bleach or any other cleaning or disinfectant agents for internal circuits. This could result in damage to
the product. Only use approved disinfectant tablets.
•
Always drain the product before disinfecting the internal water circuit. Failure to drain the product may reduce the
effectiveness of the disinfection process.
Note: Disinfection of the Altrix internal water system was validated using M. mucogenicum.
Tools Required:
•
2 gallons (7.6 L) of sterile distilled water or distilled water that has been passed through a filter less than or equal to
0.22 microns
•
Personal protection equipment (PPE) as recommended by the disinfectant manufacturer's instructions
•
Soft, lint free cloth (2 or more)
•
2 BruClean TbC tablets (active ingredient is sodium dichloroisocyanturate (NaDCC ); 1874 ppm disinfectant
solution when dissolved in one gallon of water) or equivalent
•
Service tool adapter hose (8001-999-017) for Colder style connector hoses
•
Floor drain
See
Product illustration on page 1-11
Draining the internal water circuit and hoses for disinfection
1.
Unplug the power cord from the wall outlet.
www.stryker.com
Disinfecting
for clarification of product component names and locations.
8001-109-005 REV B