Specifications (Continued)
EN
Controller
Heater capacity, maximum
Circulating fluid
Battery
Alarm tone range
Water flow rate in each hose
port
Refrigerant type
Power cord length
Clinical thermometer
Equipment Class
Note: The controller takes approximately 9 minutes to heat from 23.0±2° C (73.4° F) to 37.0° C (98.6° F) when not
connected to a patient. Time will vary when connected to a patient.
Stryker reserves the right to change specifications without notice.
For more information about thermal transfer devices, cables, or probes, see the manufacturer's instructions for use.
Environmental
conditions
Ambient temperature
Relative humidity (non-
condensing)
Atmospheric pressure
1-10
Introduction
500 watts
Sterile distilled water
9V Lithium
75 - 85 dBA per standard IEC 60601-1-8
Typical 1.2 lpm
R134a
14 to 15 feet
Direct mode
Class I
Rated for continuous operation
Operation
Storage
89.9°F
(32.2°C)
59°F
(15°C)
75 %
30 %
1060 hPa
700 hPa
8001-109-005 REV B
4.2 - 4.5 meters
Transportation
140 °F
(60 °C)
-40°F
(-40°C)
95 %
10 %
1060 hPa
500 hPa
140 °F
(60 °C)
-20°F
(-29°C)
95 %
25 %
Not applicable