Disinfecting the internal water circuit and hoses (Continued)
EN
5.
Disconnect the bottom hose from the bottom right port
6.
Connect the bottom hose end to the hydraulic connector in the lid of the reservoir
7.
Plug the power cord into a wall outlet.
8.
Press and hold the
9.
Tap the Manual mode icon.
10. Tap Confirm.
11. Set the water target temperature to 25.0° C (77.0° F).
12. Allow the controller to run for 2 minutes to purge air from the system.
13. Tap Confirm.
14. Run the controller for 20 minutes.
15. After 22 minutes, turn the controller off by pressing and holding the Stand-by button for two seconds.
16. Unplug the power cord from the wall outlet.
17. Place the controller over a floor drain.
18. Remove the reservoir. Pull forward at an angle, and lift out the reservoir.
19. Remove the bottom hose end from the hydraulic connector adapter in the reservoir lid by pushing down on the
collar.
1-44
Disinfecting
(Figure 1-30 on page
Figure 1-30: Disconnected hose
Figure 1-31: Bottom hose end in the lid of the reservoir
Stand-by button.
8001-109-005 REV B
1-44).
(Figure 1-31 on page
1-44).