Check patient probe alarm (Continued)
EN
Primary patient temperature changes by more than 1.0° C within two minutes.
Notes
•
The product will deactivate the heat exchange and keep the pump activated as requested by the Active Therapy.
•
The audio paused timeout is 5 minutes for the check patient probe alarm.
Patient probe malfunction alarm
This alarm notifies the operator that the probe is not providing information to the product during an active therapy.
Alarm generation:
When the primary patient probe is in a shorted, opened condition, or out of range for more than 30 seconds, the product
will display the Patient Probe Malfunction Alarm.
Notes
•
The product will deactivate the heat exchange and keep the pump activated as requested by the Active Therapy.
•
The audio paused timeout is 5 minutes for the patient probe malfunction alarm.
Patient probe disconnect alarm
This alarm notifies the operator that the probe is not providing information to the product during an active therapy.
Alarm generation:
When the adapter cable for the primary probe is removed and the reading of the Primary Patient probe is out of range
for more than 30 seconds, the product displays the Patient Probe Disconnected alarm.
Note: The audio paused timeout is 5 minutes for the patient probe disconnected alarm.
Patient temperature deviation medium alarm
This alarm notifies the operator that the patient is not responding as expected in the active therapy.
Alarm generation:
The product will display the Patient Temperature Deviation Medium Alarm, if after the initial achievement of the current
patient temperature target, the actual primary patient temperature becomes more than 1.0° C above or below the
Current Target Temperature.
Note: The audio paused timeout is 30 minutes for the patient temperature deviation medium alarm.
Patient temperature deviation low alarm
This alarm notifies the operator that the patient is not responding as expected in the active therapy.
Alarm generation:
The product will display the Patient Temperature Deviation Low Alarm, if after the initial achievement of the current
patient temperature target, the actual primary patient temperature becomes more than 0.5° C above or below the
Current Target Temperature.
Note: The low priority audible alarm is repeated every 30 minutes.
1-54
Alarm conditions
8001-109-005 REV B