Herunterladen Inhalt Inhalt Diese Seite drucken

English; Starting Monitor Mode; Switching Modes; Pausing And Resuming Therapy - Stryker Altrix Handbuch

Precision temperature management system
Vorschau ausblenden Andere Handbücher für Altrix:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 141
Starting Manual mode (Continued)
A temperature probe is not required when operating in Manual mode.
After the water target temperature is achieved, water temperature is controlled within +/-0.3° C.

Starting Monitor mode

In Monitor mode, no therapy will be delivered, only a display of the current patient temperature.
To start Monitor mode:
1.
Connect the reusable adapter cable to port A or port B on the controller. Make sure that the probe is fully seated.
2.
Apply the sensing end of patient probe to patient based on hospital protocol. Secure the patient probe to reduce the
risk of accidental dislodgment.
3.
Tap the Monitor button.
4.
Connect the patient temperature probe to the end of the reusable adapter cable. See
temperature probe on page
Note: If the product senses a patient probe temperature below 36.0° C (96.8° F) or above 38.0° C (104° F), the
normothermia alarm displays and an audible alarm sounds.
5.
Tap Confirm. The screen will display the current patient temperature.

Switching modes

Tap Edit and select a different therapy mode.

Pausing and resuming therapy

To pause therapy, press and hold the Pause Therapy button for two seconds.
To resume therapy, press and hold the Pause Therapy button for two seconds.

Displaying the data storage

The system gathers data at 60 second intervals and is limited to 90 minutes of storage. The graphical display defaults to
show data for all four variables in the Manual, Monitor, and Automatic modes.
www.stryker.com
Operation
1-23.
8001-109-005 REV B
Selecting and connecting a
1-33

English

EN

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis