Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
WARNING
Always use Stryker accessories. Only IEC 60601-1 equipment shall be hooked to the patient temperature ports.
Failure to comply with these instructions may invalidate any or all warranties and may negatively affect the products
EMC performance. This also protects the product from cardiac defibrillation.
Avoid reduction in water flow. Do not connect two or more thermal transfer devices in a series on a single port.
When you operate the product near ambient temperature limitations of 15.0° C (59.0° F) or 32.0° C (89.6° F), you
may experience a reduction in product performance.
Always turn or re-position the patient over the duration of therapy, if possible, to reduce the risk of pressure ulcers.
Follow your hospital protocol.
Always check the integrity of the patients skin and temperature according to hospital protocol when using the Altrix
system.
When using the temperature controlled Automatic therapy mode for warming (min, med, or custom), switching to
other modes, changing the target patient temperature, or changing the therapy selection may impact the overall
benefit of therapy.
CAUTION
Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in
this manual.
Do not modify the product or any components of the product. Modifying the product can cause unpredictable
operation resulting in injury to patient or operator. Modifying the product also voids its warranty.
Shock Hazard - Improper handling of the power cord may damage the power cord and cause potential shock
hazards. If damage has occurred to the power cord, immediately remove the temperature management system
from service to avoid the risk of serious injury or death. Contact the appropriate maintenance personnel.
Take special precautions regarding electromagnetic compatibility (EMC) when using medical electrical equipment
like Altrix. Install and place Altrix into service according to the EMC information located in the EMC section of this
manual. Portable and mobile RF communications equipment can affect the function of Altrix.
Shock Hazard. If the internal electrical components are exposed, because the side panel or cover are compromised,
remove the product from use.
Always make sure that the product reaches room temperature before you setup or operate the product.
Before first use, disinfect the internal water circuit.
Do not use Altrix located near or stacked with other medical equipment. If it is necessary to locate Altrix near other
medical equipment, make sure it operates as intended.
Portable RF communications equipment, including peripherals such as antenna cables and external antennas,
should be used no closer than 12 inches (30 cm) to any part of Altrix system, including cables specified by the
manufacturer.
Always apply the wheel locks to prevent unintended movement.
Avoid the use of materials of good thermal conductivity, such as water, gel, or similar substances, with the Altrix
system not powered on. This can decrease the temperature of the body of a patient.
Do not apply thermal transfer devices to patients with ischemic limbs. This may result in harm to the patient.
Do not use this product if the patient has a transdermal medication (patch) as this can result in increased drug
delivery.
Always pre-fill the thermal transfer devices with sterile distilled water before you apply it to the patient. This is to
reduce the risk of pressure ulcers.
Electric shock. This equipment must only be connected to a supply mains with protective earth.
www.stryker.com

Summary of safety precautions

8001-109-005 REV B