Disinfecting the internal water circuit and hoses (Continued)
20. Empty water from the reservoir, dispose of the water per hospital protocol.
Note: Do not rinse the reservoir.
21. Pull up on the controller drain plug
22. Make sure that all 3 hoses remain raised above the connection ports for draining.
23. Allow the controller and hoses to drain for a minimum of two minutes.
24. Push down on the controller drain plug to close the drain.
25. When the controller and hoses are drained, continue to
Rinsing the internal water circuit and hoses
1.
Using appropriate measuring equipment, fill the empty reservoir with 1 gallon (3.8 L) of sterile distilled water or
distilled water that has been passed through a filter less than or equal to 0.22 microns.
2.
Place the reservoir into the controller.
3.
Connect the bottom hose end to the hydraulic connector in the lid of the reservoir.
4.
Plug the power cord into a wall outlet.
5.
Press and hold the
www.stryker.com
Disinfecting
(Figure 1-32 on page
A
Figure 1-32: Drain plug
Rinsing the internal water circuit and hoses on page
Figure 1-33: Bottom hose end in the lid of the reservoir
Stand-by button.
8001-109-005 REV B
1-45) to open the drain.
English
EN
1-45.
1-45