Product alarms (Continued)

Alarm priority and description

Priority alarm
Audible reminder
Medium
Repetitive burst of 3 beeps upon first
occurrence and every 25 seconds until the
condition is resolved.
If you pause the audio when the timer expires,
the audio indication will resume.
Low
Single burst of 2 beeps upon first occurrence.
Audio pause
The audio pause button breathes as a reminder
and resets all of the active alarms.
The audible alarm resumes as soon as a new
alarm is raised or the last Medium audio paused
timer has expired.
You can press the audio pause button at any
time during the alarm condition, as many times
as necessary.
Icon, yellow
Name
Water
temperature
deviation
No water
No water flow
Check water flow Medium, 10
Check patient
probe (A or B)
www.stryker.com
Introduction
Alarm priority
Message
and delay
Medium, 5
Water
minute delay
temperature is
±0.8° C (1.4° F)
outside of target
temperature
Medium, 60
No water
second delay
Medium, 20
No flow detected
second delay
Reduced flow
minute delay
detected
Medium
Abnormal
change in patient
temperature
8001-109-005 REV B
Icon
When a medium priority alarms, the icon will
breathe yellow to indicate there is an alarm.
The icon will continue to breathe yellow until
the alarm is resolved.
The icon for the low priority alarm is solid
yellow. The icon will remain solid yellow until
the alarm condition is resolved.
Pausing the alarm will not stop the icon from
breathing yellow or being solid yellow.
Therapy
Check
interrupted
No
Temporary condition
upon startup,
addition of thermal
transfer device, or
addition of water
Yes
Check for leaks
Add a minimum of 2
liters of water
Yes
Check for leaks and
obstructions at
connections, hoses,
and thermal transfer
devices
No
Tap Confirm, if the
water port was
removed
intentionally
Check for leaks and
obstructions at
connections, hoses,
and thermal transfer
devices
Yes
Check probe
condition, location,
and connections