Starting Automatic therapy mode (Continued)
2.
See
Filling a thermal transfer device on page
3.
Apply the thermal transfer device to the patient.
4.
Connect the reusable adapter cable to port A or port B on the product. Make sure that the probe is fully seated.
5.
Apply the sensing end of a patient probe to the patient based on your hospital protocol and secure the product to
reduce the risk of accidental dislodgment.
6.
Connect the patient temperature probe to the reusable adapter cable. See
probe on page
1-23.
7.
Tap to Confirm the current patient temperature.
8.
Tap the Automatic therapy mode button.
9.
Select the target patient temperature.
10. See
Setting or editing the cooling rates on page 1-31
Notes
•
The controller determines Warming or Cooling therapy based on the selected water target temperature and
the current water temperature.
•
Do not place additional heat sources between the patient and thermal transfer device.
•
After the patient target temperature is achieved, patient temperature is controlled within +/-0.3° C.
•
If the patient temperature is not within 0.5° C of the current target temperature, the yellow patient icon will flash
and the patient temperature deviation alarm will sound. This occurs after the initial patient target temperature is
achieved.
Setting or editing the cooling rates
Setting cooling rates is for Automatic Mode only.
1.
To set the cooling temperature, highlight your choice of cooling rates.
Select a cooling rate
2.
Tap Confirm.
3.
Tap the Edit button to make changes.
Setting or editing the warming rates
Setting warming rates is for Automatic Mode only.
1.
To set the warming temperature, highlight your choice of warming rates.
www.stryker.com
Operation
1-29.
or
Setting or editing the warming rates on page 1-31
Description
Maximum: approaches patient target temperature as fast as possible
Medium: water is cooled to target, with a max of 15.0° C difference between the
patient and the water temperature
Minimum: water is cooled to target, with a max of 10.0° C difference between the
patient and the water temperature