Powering on the product (Continued)
CAUTION (CONTINUED)
Explosion risk. This product is not suitable for use in the presence of flammable anesthetic mixture with air or with
oxygen or nitrous oxide other than nasal or mask type.
Do not place cables, hoses, or power cord in walkways to avoid the risk of trip hazards.
Do not use three or more adult Mul-T-Blanket products at the same time to avoid the risk of water overflow when
you power off the controller.
To start the product:
1.
Plug the power cord into a wall outlet.
2.
Tap the Stand-by button to start the product.
3.
If you are going into Automatic mode or Monitor mode, see
manual mode, got to the next step.
4.
See
Removing and replacing the reservoir on page
5.
See
Filling the reservoir with sterile distilled water on page
6.
Connect up to three thermal transfer devices (with the exception of adult Mul-T-Blankets) to dedicated adapter
hoses and ports.
7.
Open the clamps on the connector hose and the thermal transfer devices to provide proper water flow.
8.
See
Filling a thermal transfer device on page
9.
See
Selecting a therapy mode on page
10. Make sure that the desired port configuration is maintained and that water is flowing through the thermal transfer
devices.
WARNING
Always turn or re-position the patient over the duration of therapy, if possible, to reduce the risk of pressure ulcers.
Follow your hospital protocol.

Removing and replacing the reservoir

The removable reservoir enables you to fill or drain the reservoir away from the controller without interrupting therapy.
You will need to have the reservoir installed before starting a therapy.
CAUTION
Do not place your fingers in between the reservoir and the sides of the controller, to avoid the risk of pinching your
fingers.
www.stryker.com
Operation
Selecting and setting the primary probe on page
1-27.
1-28.
1-29.
1-30.
8001-109-005 REV B
English
EN
1-29. If
1-27