Herunterladen Inhalt Inhalt Diese Seite drucken

Disinfecting The Internal Water Circuit And Hoses - Stryker Altrix Handbuch

Precision temperature management system
Vorschau ausblenden Andere Handbücher für Altrix:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 141
Draining the internal water circuit and hoses for disinfection (Continued)
Figure 1-28: Clik-Tite hose ends are closed and clamps are open
6.
To fully drain the hoses, raise all the hoses
Note: For best performance, hang the hoses to keep them raised. Do not lower the hoses until you have completed
the disinfection and rinsing process.
7.
Allow the controller and hoses to drain for a minimum of two minutes.
8.
Push down on the drain plug to close the drain.

Disinfecting the internal water circuit and hoses

1.
Use personal protection equipment as recommended by the BruClean TbC disinfectant manufacturer's instructions
for use.
2.
Put 2 BruClean TbC tablets into the reservoir.
3.
Using appropriate measuring equipment, fill the empty reservoir with 1 gallon (3.8 L) of sterile distilled water.
Note: Always allow the disinfectant tablets to completely dissolve before starting the 20 minute disinfection
cycle.
4.
Place the reservoir into the controller.
www.stryker.com
Disinfecting
B
(Figure 1-29 on page
1-43) above the connection ports on the controller.
Figure 1-29: Raise the hoses
8001-109-005 REV B
A
English
EN
1-43

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis