Removing and replacing the reservoir (Continued)
EN
To remove the reservoir, pull forward at an angle, and lift out the reservoir
1.
To replace the reservoir, align the base of the reservoir over the drain (C).
2.
Align the notch on the back of the reservoir (A) with the hook on the controller (B)
3.
Push the reservoir back into place. Make sure that the reservoir is secure to avoid water leakage.

Filling the reservoir with sterile distilled water

The removable reservoir is translucent for you to see the water levels.
CAUTION
Always use sterile distilled water or distilled water that has been passed through a filter less than or equal to 0.22
microns with this product.
Always fill the reservoir with room temperature sterile distilled water to reduce the risk of burn.
Do not overfill the reservoir to avoid the risk of water spillage and fall.
To fill the removable reservoir with sterile distilled water:
1.
See
Removing and replacing the reservoir on page
1-28
Operation
B
C
Figure 1-16: Removable reservoir
1-27.
8001-109-005 REV B
(Figure 1-16 on page
1-28).
A
(Figure 1-16 on page
1-28).