Adapting the Armrests to the Forearm Length
1.  Loosen the two set screws on the underside of the armrest 
using a 3 mm Allen wrench. On the side where the control 
panel is mounted, three set screws have to be loosened 
(see Fig. 11).
2.  Move the armrest with the control panel receiver or the 
replacement tube along the attachment rail forwards or 
backwards into the desired position.
3.  Firmly retighten all set screws.
Fig. 11
Adapting the armrests to the forearm length
C2000
6.1.3 Control Panel
CAUTION
Risk of accidents and injury due to uncontrolled moving
of the wheelchair. Turn off the control unit of the power 
wheelchair before mounting or removing the control panel 
or changing its position.
NOTICE
Cable damage. Positioning the cable incorrectly can lead 
to pinching and thus damage to the cable. The cable must 
not be attached too tightly or too loosely. Avoid bending or 
squeezing the cable.
Adapting the Control Panel to the Arm Length
The control panel is attached to a rail which is located beneath 
the armrest.
1.  Loosen the three set screws on the underside of the arm-
rest.
2.  Move the control panel with the control panel receiver 
along the attachment rail forwards or backwards into the 
desired position.