Herunterladen Inhalt Inhalt Diese Seite drucken

Incubator In The Side By Side Configuration; Side View - Stryker Power-PRO IT 6510 Bedienungs- Und Wartungshandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 175
English
INSTAllING THE AIRBORNE
Prior to installing the Airborne™ Side By Side Incubator on the model 6510 Power-PRO™ IT cot, read and understand
this manual and the manual supplied with the incubator.
WARNING
The Airborne™ Side By Side Incubator adaptor (6510-028-000) is designed to secure only Airborne™ incubators to the
model 6510 Power-PRO™ IT cot. Using this adaptor on any cot other than the model 6510 Power-PRO™ IT cot or using
any unapproved incubators on this adaptor may result in damage to the cot and/or injury to the patient or user.
To install the incubator:
1.
Push down on latch tab (A) to release latch tab (B) as shown in Figure 27.1.
2.
Pull down on latch tab (B) and open each of the latches on the four corners of the incubator.
3.
Place the Airborne™ incubator into the adaptor on the cot. Verify that all four corners of the incubator are properly
seated in the adaptor.
4.
Insert each latch into its slot on the adaptor. Push up on latch (B) to secure the latches. Verify that all four latches
are securely fastened.
Figure 27.1: Airborne™ Side By Side Incubator
WARNING
Verify that the adaptor is properly installed on the cot and the incubator is securely fastened to the adaptor prior to use.
An improperly attached adaptor or incubator may cause injury to the patient or user.
Return To Table of Contents
1-50
Cot Operation

INCUBATOR IN THE SIDE By SIDE CONFIGURATION

B
B
A
A
6510-009-005 REV A

SIDE VIEW

Figure 27.2: latch Tabs - Side View
www.stryker.com
B
B
A
A

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis