The condition of the cot is the responsibility of the owner. It is important that the cot is working properly before the
product is put into service. Have a qualified service technician use the following list and the operation instructions to
check the cot before the product is put into service.
The battery must be charged prior to checking the features and condition of the cot.
Item
Battery
The power indicator LED, located at the foot end control enclosure of the cot, is solid green when the battery is fully
charged or has adequately charged battery power.
CAUTION
When charging a battery in an ambulance, locate the charger in an enclosed cabinet and out of patient reach during
transport.
After the battery is fully charged, inspect the cot for the following points:
Item
Battery
Hydraulics
Electronic
Controls
Manual Back-up
Release
Return To Table of Contents
1-20

Product Inspection

Unpack batteries and charger
Charge battery according to SMRT Power System instructions (6500-009-101) or D
Battery System instructions
Charge spare battery (if necessary) according to SMRT Pak or D
battery instructions
Install battery into foot end enclosure - power indicator LED operates
Ensure that the battery remains firmly secured
Release and remove battery from foot end enclosure
Reinstall battery into foot end enclosure
Inspect motor mount - all fasteners secure
Check cylinder attachments at both ends - all fasteners secure
Inspect main cable - all connections secure
Inspect hoses and cylinder seal for leaks
Check power indicator LED - charged
Extend cot to raised position
Verify "jog" function operates smoothly
Lower to retracted position - cot secures in a mid-height position
(motor does not operate)
Determine and set ambulance vehicle load height
Check high speed retract
Extend cot to full height - no drift
Verify the manual back-up release handle functions properly - adjust
accordingly
With the cot empty, check the raise/lower function
With the cot loaded with a minimum of 99 pounds (45 kg), check the raise/
lower function
With the cot loaded with a minimum of 99 pounds (45 kg), check the load/