REMOVING AND REPlACING A D
WARNING
To avoid risk of electric shock, never attempt to open the battery pack for any reason. If the battery pack case
is cracked or damaged, do not insert it into the charger. Return damaged battery packs to a service center for
recycling.
Do not remove the battery when the cot is activated.
Avoid direct contact with a wet battery or battery enclosure. Contact may cause injury to the patient or operator.
To remove the battery:
1.
Press the red battery release button (A), located on the patient left side of the foot end control enclosure, to
release the battery (B) from the cot as shown in Figure 21.
2.
Slide the released battery out of the enclosure.
To reinstall or replace the battery:
1.
Align the tabs in the battery enclosure.
2.
Push the battery into the enclosure until the latch clicks into place.
The cot power indicator LED is solid GREEN if the battery is fully charged and ready.
The cot power indicator LED flashes RED if the battery needs to be recharged or replaced.
Note: Batteries slowly lose power when not on the charger.
CAUTION
Remove the battery if the cot is not going to be used for an extended period of time (more than 24 hours).
www.stryker.com
Operation Guide
WAlT
®
BATTERy
e
A
Figure 21: D
WAlT
®
Battery Removal & Replacement
e
6510-009-005 REV A
B
Return To Table of Contents
English
1-43