WARNING (CONTINUED)
Two operators must be present when the cot is occupied.
Operators must be able to lift the total weight of the patient, cot and any items on the cot.
The higher an operator must lift the cot, the more difficult it becomes to hold the weight. An operator may need
help loading the cot if he/she is too short or if the patient is too heavy to lift safely. The operator must be able to
lift the cot high enough for the cot legs to unfold completely and lock when the cot is unloaded. A shorter operator
needs to raise their arms higher to enable the undercarriage to unfold.
There must be a safety hook properly installed in the vehicle so that the bumper does not interfere with the front
legs of the base frame.
Failure to install the safety hook can cause injury to the patient or operator. Install and use the safety hook as
described on
page
1-26.
When using a standard cot fastener, do not load the cot into the vehicle with the head section retracted. Loading
the cot with the head section retracted may cause the product to tip or not engage properly in the cot fastener,
possibly causing injury to the patient or operator and/or damage to the cot.
Whenever the weight of the cot and patient is off of the wheels, the cot will automatically enter the high speed
retract mode if the retract (–) button is pressed.
Once the weight is off of the ground, the operator(s) must support the load of the patient, cot and any accessories.
Failure to support the load properly may cause injury to the patient or operator.
The one person loading and unloading procedures are for use only with an empty cot. Do not use the procedures
when loading/unloading a patient. Injury to the patient or operator could result.
Do not pull or lift on the safety bar when unloading the cot. Damage to the safety bar could result and injury to
the patient or operator could occur.
Do not press the extend (+) button until the safety bar has engaged the safety hook.
To avoid risk of electric shock, never attempt to open the battery pack for any reason. If the battery pack case
is cracked or damaged, do not insert it into the charger. Return damaged battery packs to a service center for
recycling.
Do not remove the battery when the cot is activated.
Avoid direct contact with a wet battery or battery enclosure. Contact may cause injury to the patient or operator.
To avoid injury, always verify that the head section is locked into place prior to operating the cot.
Never install or use a wheel lock on a cot with excessively worn wheels. Installing or using a wheel lock on a wheel
with less than a 6" (16 cm) diameter could compromise the holding ability of the wheel lock, possibly resulting in
injury to the patient or operator and/or damage to the cot or other equipment.
When the optional head end storage flat is being used, ensure that it does not interfere with the operation of the
retractable head section, safety bar and safety hook. Injury to the patient or operator could result.
These adaptors are intended for use only on the model 6510 Power-PRO™ IT cot. They are not intended for
installation on any other Stryker cot or on any cot from another manufacturer. Using these adaptors on any cot
other than the model 6510 Power-PRO™ IT cot may result in damage to the cot and /or injury to the patient or user.
Verify that the adaptor is properly installed on the cot and the incubator is securely fastened to the adaptor prior
to use. An improperly attached adaptor or incubator may cause injury to the patient or user.
The Airborne™ Side By Side Incubator adaptor (6510-028-000) is designed to secure only Airborne™ incubators
to the model 6510 Power-PRO™ IT cot. Using this adaptor on any cot other than the model 6510 Power-PRO™
IT cot or using any unapproved incubators on this adaptor may result in damage to the cot and/or injury to the
patient or user.
The Drager
Incubator adaptor (6510-029-000) is designed to secure only Drager
®
PRO™ IT cot. Using this adaptor on any cot other than the model 6510 Power-PRO™ IT cot or using any
unapproved incubators on this adaptor may result in damage to the cot and/or injury to the patient or user.
Stryker is not responsible for specifications changes to the Drager
The Airborne™ Stackable adaptor (6510-027-000) adaptor is designed to secure only an Airborne Stackable to the
model 6510 Power-PRO™ IT cot. Using this adaptor on any cot other than the model 6510 or using any unapproved
incubators or stackables on this adaptor may result in damage to the cot and/or injury to the patient or user.
Verify that the adaptor is properly installed on the cot and the oxygen module is securely fastened to the adaptor
prior to use. An improperly attached adaptor or oxygen module may cause injury to the patient or user.
www.stryker.com
Summary of Safety Precautions
6510-009-005 REV A
incubators to the 6510 Power-
®
(or Air-Shields
Series) incubators.
®
®
Return To Table of Contents
English
1-13