B
To use Walk Away mode - Connect the AC power adapter to the Analyzer and a power source
Step 6B: Starting Walk Away mode
• Turn the Analyzer on by pressing the blue power button once.
• When the display window reads: "INSERT TEST DEVICE OR DOUBLE-CLICK FOR
WALK AWAY MODE,"
– Double-click the blue power button.
Step 7B: Adding the specimen:
• When the display window reads "ADD SPECIMEN TO TEST DEVICE AND
INSERT IMMEDIATELY":
– Invert the tube, holding it vertically (approximately one inch above the BD Veritor
System RSV device sample well).
– Gently squeeze the ridged portion of the tube, allowing three (3) drops of the
processed specimen to dispense into the sample well of a labeled BD Veritor System
RSV device.
NOTE: Squeezing the tube close to the tip may cause leakage.
CAUTION: A countdown timer displays the time remaining for test insertion. Walk Away
mode must be activated again when this timer expires. Confirm timer is visible and
Walk Away mode is activated before inserting test device.
Step 8B: Start the development and reading sequence
• Insert the test device into the slot on the right side of the Analyzer.
The test device must remain horizontal to prevent spilling the specimen out of the
sample well.
• "DO NOT DISTURB TEST IN PROGRESS" appears in the display window. Automatic
timing of the assay development, image processing and result analysis begins.
• The display window shows the remaining analysis time.
Do not touch the Analyzer or remove the test device during this process. Doing so will
abort the assay analysis.
Step 9B: Record the Result
• When analysis is complete, the test result appears in the display window. Record the result and discard the test device
appropriately.
ATTENTION: TEST Results are NOT maintained in the display window when the device is removed or if the
Analyzer is left unattended for more than 60 minutes (when AC power adapter is connected).
Using the BD Veritor Plus Analyzer In "Walk Away" mode:
with no bar code scanning module installed
7