Voice control package

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 115
1
�������� ��� ��������
Please read this manual and follow its instructions carefully. The words warning, caution,
and note carry special meanings and should be carefully reviewed:
W W a a r r n n i i n n g g : : Indicates measures to avoid potential serious injury to the user and the
patient and/or damage to this device.
C C a a u u t t i i o o n n : : Indicates risks to the equipment. Failure to follow cautions may result in
product damage.
N N o o t t e e : : Provides special information to clarify instructions or present additional useful
information.
1.
Federal law (United States of America) restricts this device to use by, or on order of, a
physician or qualified health care professional who have a complete knowledge of the
use of this equipment and the devices to which it can be connected.
2.
Read this operating manual thoroughly and be familiar with its contents prior to setting
up or using this equipment.
3.
Read and be familiar with the operating manuals for all connected devices before using
them with the SDC3.
4.
Carefully unpack the contents of this package and check for any damage that may have
occurred during shipment.
5.
Test this equipment prior to using it in a surgical procedure. This unit was fully tested at
the factory before shipment; however, it must be tested for proper function in the
environment and setup configuration in which it will be used.
6.
Install this device in an operating room that complies with all applicable IEC, CEC, and
NEC requirements for safety of electrical devices.
7.
Do not sterilize the Jabra Wireless Headset or the Audio Technica Headset/microphone
system as product damage may result. Follow provided instructions for safely placing
the headset prior to a surgical procedure.
8.
Ensure that all equipment connected to the SDC3 is properly grounded (and isolated, if
required).
9.
To avoid risk of fire, do not use this device in the presence of flammable anesthetics.
10. Pay close attention to the care and cleaning instructions in this manual. Failure to follow
these instructions may result in product damage.
11. Attempt no internal repairs or adjustments not specifically detailed in this operating
manual. Refer any readjustments, modifications, and/or repairs to Stryker Endoscopy or
its authorized representatives.
12. Do not modify this equipment without authorization of the manufacturer.
13. This operating manual should be used in conjunction with the SDC3 base manual.
The Stryker Standard Warranty and Return Policy (1000401175) is void if any of these
warnings or cautions are disregarded.