Maintenance
Maintenance, Inspection, and Testing
• Inspect the device on a continual basis. If a problem is observed
or suspected, the device should be returned for evaluation and/or
repair.
• Inspect all components for cleanliness. If fluid or tissue buildup is
present, repeat the above cleaning and disinfection procedures.
Storage
Store the coupler in a dry, clean, and dust-free environment at room
temperatures.
Expected Service Life
The Precision AC Coupler has an expected service life of 2 years,
except with repeated use of an automated washer the expected
service life is 1 year.
Note: If the coupler is cleaned with the automated cleaning method,
steam sterilization is recommended. Repeated automated cleaning
when sterilizing with Sterrad or V-Pro can degrade the product's
cosmetic appearance.
When the device has reached the end of its service life, dispose of it
according to local laws and hospital practices.
Technical Specifications
Transportation and
Storage Conditions
Operating Conditions
EN-12
Temperature: -18–60°C
Relative Humidity: 15–90%
Temperature: 10–40°C
Relative Humidity: 25–75%