check if the illumination can be controlled.
• The handset cannot be turned on: Confirm that the battery is charged.
• The battery operating time is too short: Check the battery is in good
condition.
4.5 Information about the device
The RCS-100 is a portable and multifunctional electronic imaging dia-
gnostic system. It consists of a handset (5.0 inches, 720 p, multi-touch
display, 3.6 V 2,600 mAh rechargeable battery, storage capacity for 1,000
photos) and four interchangeable lenses (otoscopes, dermatoscope, ge-
neral lens, adapter for connecting endoscopes).
5. Technical specifications
Handset:
Medical device:
Medical device for image reproduction and
power supply for modules
Protection class/
electrical protection:
Protection class II
Class II isolation equipment
Model:
RCS-100 Riester Camera System
Plug-in power
supply/power supply
Input: 100 V-240 VAC / 50-60 Hz / 0.3 A
Output: 5 VDC / 2 A
Classification:
Application part type B
Size and weight:
Size: 225 mm × 135 mm × 45 mm
Weight: 292.0 g
LCD screen:
5.0 inch touch display
(110.7 mm × 62.3 mm), 1280 × 720 p
Focus:
Automatic/manual
Photo/video format :
Photo: JPEG, video: MP4
Battery:
33.6 V, 3.7 V 2,600 mAH 18650 Li recharge-
able battery
Video:
3.5 hours (fully charged battery with
25°C ambient temperature)
Charging adapter:
Input 100-240 V, 50-60 Hz, 0.3 A
Output DC 5 V / 2 A
CMOS pixel density
8 M
USB
OTG and type C USB
RAM
2G LPDDR3
ROM
16G
Memory extension (OPT) 16G Wifi-SD card
Otoscope:
Size and weight:
Size: 73.5 mm × 40.0 mm × 40.0 mm
Weight: 96 g
F/
2.9
Max. object distance:
15 mm, at maximum object distance FOV-
Diameter:
15 mm
Object height:
10 mm (Ø 4 speculum)
Depth of field:
10 mm
Illumination source:
LED with natural light
LED colour temperature:
4000K
Magnification on the display:6 x
Dermatoscope:
Size and weight:
Size: 62.1 mm × 62.1 mm × 36.0 mm
Weight: 108.5 g
F/
2.2
Polarisation:
Dermis/epidermis
Object distance:
0 mm
Field of view:
30 mm Ø
Optical magnification:
2.5x (from the optics)
Digital enlargement:
4x (additional)
Screen magnification:
2 x - 8 x
Illumination source:
LED with natural light
LED colour temperature:
4000K
General lens
Size and weight:
Size: 60.5 mm × 60.5 mm × 19.0 mm
Weight: 66.4 g
F/
2.0
Optical magnification:
2x
Digital enlargement:
4x
Field angle:
78º
Object distance:
30 mm to 4 m
Illumination source:
LED with natural light
LED colour temperature:
5500K
Operating conditions:
For indoor use only
Ambient temperature:
10°C / 50°F to 40°C / 104°F with a relative
humidity of 15% to 90% (non-condensing)
Storage conditions:
0°C / 32°F to 45°C / 113°F with a relative
humidity of 15% to 95% (non-condensing)
Air pressure:
700 – 1060 hPa
LA 4° adapter
Size and weight
Size: length = 51.5 mm, diameter = 54 mm
Weight: 114 g
Field angle:
4°
Object distance:
2.6 mm - 2.9 mm
Illumination source:
None
Operating conditions:
For indoor use only
Ambient temperature:
10°C / 50°F to 40°C / 104°F with a relative
humidity of 15% to 90% (non-condensing)
Storage conditions:
0°C / 32°F to 45°C / 113°F with a relative
humidity of 15% to 95% (non-condensing)
Air pressure:
700 – 1060 hPa
LA 9° adapter
Size and weight:
Size: length = 51.5 mm, diameter = 54 mm
Weight: 114 g
Field angle:
9°
Object distance:
2.6 mm - 2.9 mm
Illumination source
None
Operating conditions:
For indoor use only
Ambient temperature:
10°C / 50°F to 40°C / 104°F with a relative
humidity of 15% to 90% (non-condensing)
Storage conditions:
0°C / 32°F to 45°C / 113°F with a relative
humidity of 15% to 95% (non-condensing)
Air pressure:
700 – 1060 hPa
6. Spare parts and accessories
Item no.:
13270
Dermatoscope attachment for RCS-100 camera
13271
General lens for RCS-100 camera
13272
Otoscope attachment for RCS-100 camera
13277
LA 4° adapter for RCS-100 camera
13278
LA 9° adapter for RCS-100 camera
13274
Black case for RCS-100 complete with inserts
7. Maintenance/accuracy check/calibration/applied standards
The instruments and their accessories require no special maintenance.
If an instrument needs to be tested for any reason, please send it to us
or an authorized Riester dealer in your area, the details of which we will
provide you with upon request.
7.1 Applied standards
IEC 60601-1
IEC 60601-1-2
8. Disposal
Caution!
The used medical device must be disposed of in accordance with current
medical practices or local regulations on the disposal of infectious bio-
logical medical waste.
Batteries and electrical/electronic devices may not be treated as
domestic waste and must be disposed of in accordance with local re-
gulations.
If you have any questions about the disposal of products, please
contact the manufacturer or their representative.
9. Electromagnetic compatibility
The instrument satisfies the requirements for electromagnetic com-
patibility. Please note that under the influence of unfavourable field
strengths, e.g. when operating cell phones or radiological instruments,
malfunctions cannot be ruled out.
In accordance with the requirements of IEC60601-1-2:2014, the electro-
magnetic compatibility of this device has been verified in a test.
25