1.3.4.9
Alarms and Indicators
The
Datascope ACCUTORR
provides high and low alarrn limits for systolic pressure,
diastolic pressure, mean pressure, heart rate, and SaOz*,
A. Setting Alarm Limits
1. Press ALARMS SELECT (27) to activate the ALARM SET MODE and to choose a
desired alarm parameter. The alarm limits for each parameter are displayed in a win-
dow superimposed on the trend display. Continue to press ALARMS SELECT (27) to
cycle through the available alarm parameters. Thirty seconds is allowed to begin set-
ting the alarm limits once you are in the ALARM SET MODE.
NOTE:
Pressing and holding the ALARMS SELECT key for 3 seconds will cancel all
of the alarm limits. All limits will be set to OFF except for SaO2 Low, which will be set
to 85.
2. Press HI (26) until
a
the desired numerical
1 [MED
|
MIN
|
value for the high alarm
Tr
==
|
limit displays. See Sec-
M
tion 1.3.4.8, B, Alarm
Limits, for the available
high alarm limits.
3. Press LO (28) until
the desired numerical
value for the low alarm
limit displays. See Sec-
tion 1.3.4.8, B, Alarm
Limits, for the available
low alarm limits.
4. Press VOLUME (25) untii the desired alarm tone level for systolic pressure, dias-
tolic pressure, mean pressure, and heart rate is achieved (low, medium, or high).
The OFF selection may be-disabled or activated in the User Configuration Mode, see
section 1.3.4.12.
lf the alarm OFF selection is activated, a rectangular marker will ap-
pear in the upper left corner of the VOLUME window.
If
[E
MED
[에세이
MED
| Dsablen
the alarm OFF is disabled, then no marker will appear.
5. lf you are using an ACCUTORR 3 SAT or an ACCUTORR 4 SAT, press BEEP
VOLUME (6) until the desired alarm tone level for SaOe is achieved (low, etc...).
6. If desired, press MUTE (5) to silence the alarm tone for the current NIBP measure-
ment. The SaO2 alarm tone will be muted for two minutes.
Notes:
1. An NIBP measurement can be taken while setting the alarm limits. 2. You
can not set the alarm limits while printing data. 3. All alarms disable while being set.
* if equipped
Revised 08/24/90
1-27