1.5
Specifications
SaO2
Accuracy (%)
Mean
Measurement Range
# of digits
Difference
Adult
Neonate
SaO2
3
70 to 100% <+/-2% Sa02
Oto 100
0 το 100 (%)
60 to 70%
< +/-4% Sa02
<60%, unspecified
NIBP
Measurement Range
# of dioîts
Adult
_Neonate
Systolic Pressure
3
50 to 235
30 to 200 (mmHg)
Diastolic Pressure
3
80 to 200
10to 150 (mmHg)
Mean Pressure
3
7010220
2010 170 (mmHg)
Heart Rate
3
30 to 220
30to 250 (bpm)
Accuracy
・
Meets AAMI standard for automatic sphygmomanometers.
'
(For all available size cuffs except the thigh cuff.)
Measurement Cycle:
Less than 30 seconds average at 72 bpm without
motion artifact or arrhythmia.
Cycle time is affected by
arm size and wrapping technique, which determine cuff
capacity.
Initial Cuff Inflation (Auto Mode)
Adult/Pediatric:
180 +/-15 mmHg
Neonate:
120 +/-15 mmHg
Cuff Pressure Range
Adult/Pediatric:
0 to 270 mmHg
Neonate:
0 to 230 mmHg
ーー
Maximum Cuff Pressure:
330 mmHg (will not exceed 10 mmHg for longer
(Adult Mode)
than five minutes)
Hose Connections:
LUER-Lock Connecter.