EXCHANGE PROGRAM
Datascope offers an exchange program for certain assemblies in the instrument.
In
many cases, replacement of the complete assembly will result in the most expedient
repairs. See section 3.4 for details concerning the exchange program.
2.4 Equipment and Special Tools Required
Description
ification
DVM
Standard Mercury Column
0-300mmtig
Dummy Cuff (693)
0138-00-0001-01
Oscilloscope
Safety Analyzer
Dempsey Model or equivalent
Finger Sensor Probe
ACCUSAT Compatable
Extender Boards:
Interface Board
0670-00-0398
Power Supply and
Sa02 Analog Board
0670-00-0399
2.5 Disassembly Instructions
The ACCUTORR 3/4 design consists of sub-assemblies, individual PC boards and a
user-replaceable software cartridge.
The face plate has the ability to be oriented in one of two possible positions. When
removing the face plate assembly be sure to note the position of the mounting bracket
for later reassembly.
Before disassembling the unit, perform the following:
A. Power down the ACCUTORR and remove the power cable.
B. Disconnect all cables from the rear panel.
C. ifthe battery option is installed, disconnect and remove the battery.
D. Remove the Datasette cartridge from the back of the unit.
Note: The numbers in parentheses (} refer to the isometric drawings.