The following messages may be displayed on the LCD during a NIBP measurement
cycle. The messages are divided into four major categories: NIBP Measurement,
Qualifying NIBP Measurement, Monitor Operation, and SaOe,
A.
NIBP Measurement Messages
Message
Beason
Response
RETRY-MOTION
ARTIFACT
Too much motion artifact.
Unit begins a retry
measurement after venting
the cuff for 5 seconds.
RETRY-PUMP
HIGHER
Cuff not inflated enough to
Unit begins a retry
.
measure patient's pressure.
measurement after venting
the cuff for 5 seconds.
CUFF
OVERPRESSURE
Cuff pressure exceedsthe
Cuff vents and remains
preset trip point (330mmHg). vented. Cycle power to
reset this condition
The alarm tone sounds
for 5 seconds.
RETRY
Measurement cycle too
Unit begins a retry
.
long or no pulsations
measurement after venting
detected.
the cuff for 5 seconds.
UNABLE
TO
MEASURE
Unit cannot successfully
Alarm tone sounds
complete the NIBP
for five seconds. Correct
measurement after four
problem.
attempts.
B. Monitor Operation Messages
The following messages pertain to the operation of the monitor.
Message
Reason
Response
LOW
BATTERY.
Battery requires
Recharge/Replace
recharging.
battery. See Section 4.3
in the Operating Instructions.
INTERNAL
TEST
IN
Self test is being
Wait for completion
PROGRESS
performed.
of self-test.
See Section 1.3.2.
RAM
TEST
FAILED
Message appears until
Power cycle unit.
unit is turned
off. Self test
was not successful.
ROM
TEST
FAILED
Message appears until
Power cycle unit.
unit is turned
off. Self test
was not successful.
Revised 03/13/94