The ACCUTORR will respond to a command (except "G") by displaying the command
character, a colon then a sequence of 3 digit numbers separated by a comma.
The
last number will be followed by a carriage return and line feed. The number of 3 digit
fields and the meaning of each field is dependent on the particular command.
Ifa field
is not in use, it will be represented by"--". An example would look exactly like this (all
characters are printable):
P:--,85,120, 45, 80,000,010,000, 5
Command Responses:
The following is returned by the "Prompt request" command ("C")
>
The following is returned by the "Show alarm limits" command ("L")
L:aaa,bbb,ccc,ddd,eee, fif,ggg,hhh, iii, jj
The fields are as follows:
: Sa02 high limit
: SaO2 low limit
: Pulse rate high limit
: Pulse rate low limit
: Systolic high limit
Systolic low limit
: Mean blood pressure high limit
: Mean blood pressure low limit
Diastolic high limit
Diastolic low limit
The following is returned by the "Show patient data" command ("P")
P:aaa;bbb,ccc,ddd,eee,
fff,ggg, hhh, iii
The fields are as follows:
: Sa02
: Pulse rate
: Systolic blood pressure
: Mean blood pressure
: Diastolic blood pressure
Status Byte O (see next page)
: Status Byte 1 (see next page)
: Status Byte 2 (see next page)
i: Elapsed time (minutes) since last NIBP measurement
Revised 08/24/90
:
1-53