B. Clearing Trend
1. Press and hold DURATION (16) for three seconds.
The LCD advisory message reads, TREND CLEARED.
(This message supercedes
any other message.)
The alarm tone sounds at the high volume for three seconds.
2. The monitor automatically returns to a normal operational state once the trend
memory is cleared and the alarm tone stops.
NOTE:
Clear trend after each patient.
1.3.4.11
Printer
The DATASCOPE ACCUTORR 3 and the ACCUTORR 3 SAT provide a permanent
record of a patient's: systolic pressures, diastolic pressures, mean pressures, heart
rates, and SaO2.*
A. Operation of Printer in the NIBP Trend List Mode.
1. If necessary, press PAPER FEED (15) to advance the paper in the printer.
2. Press RECORD (11) to enter into a continuous list recording mode.
The entire
trend list is printed.
Subsequent measurements are also recorded. The LCD display
shows the status of the recorder, CON for continuous and blank for standby.
3. Press RECORD a second time to stop the recording feature.
B. Operation of Printer in the Graphic Trend or Sa02
Waveform Mode
1. lf necessasry, press PAPER FEED (15) to advance the paper in the printer.
2. Press RECORD (11) to produce a permanent record of the graphic display on the
LCD trend area. The LCD graphic display is frozen until the recording is completed.
3. Press RECORD at any time during printing to abort the process of printing.
NOTE:
When tearing off recording, tear at an angle to avoid pulling out excess paper.
See Section 4.4, in the Operating Instructions, for Paper Replacement instructions.
* if equipped