Detailed Operating Instructions
3.1 Setting-Up
1. Set the rear panel MAINS POWER SWITCH (84) OFF.
2. Check the serial number label (38) and voltage indicated. Confirm proper voltage
configuration. If the monitor is not wired correctly, contact a Datascope Service Repre-
sentative or qualified hospital personnel.
3. if interfacing with other compatible Datascope instruments, attach the interface
cable (P/N 0012-00-0271) between the rear panel INTERFACE CONNECTOR
(37)
and the corresponding interface connector on the peripheral instrument(s).
4. If additional communications capabilities are required, attach the RS232 Interface
Cable to the rear panel
RS232 Connector,* Data Out (36) and the corresponding in-
terface connector on the peripheral instrument(s).
. 5, Attach the AC power cord into the Datascope ACCUTORR's rear panel INPUT
MODULE (33) and into a grounded (3-prong) Hospital Grade AC receptacle. Do not
use an adaptor to defeat the U-ground.
WARNING
When attached to other products insure that the total
chassis leakage currents of all units (total) do not exceed
100 ua.
3.2
Turning Power On
1. Place the rear panel MAINS POWER SWITCH (34) to ON.
A green LED, MAINS ON
(21), should illuminate on the front panel.
2. Press the front panel ON/ST'BY switch (14) to ON.
The message "INTERNAL TEST IN PROGRESS" cisplays on the LCD advisory panel.
The following items are checked during the first 10 seconds of monitor operation
(initial power up):
e
Internal Self-Check
e
All Front Panel LEDs
e
The LCD Advisory Panel
e
The Alarm Tones
NOTE:
To insure proper function, watch for all front pane! LEDs and the LCD
Advisory Panel to light up.
~
The monitor is ready for use when the message is erased from the LCD.
* Before using the communication connectors, contact authorized service personnel
for assistance. Detailed Operating Instructions