Herunterladen Inhalt Inhalt Diese Seite drucken

General Information; Disassembling The Product Before Carrying Out The Reprocessing Procedure; Preparations At The Place Of Use; Preparation Before Cleaning - B. Braun Aesculap Spine Quintex SC420R Gebrauchsanweisung/Technische Beschreibung

Anteriores zervikales plattensystem – instrumente
Vorschau ausblenden Andere Handbücher für Aesculap Spine Quintex SC420R:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 24
Aesculap
Quintex® anterior cervical plate system - Instruments
7.3

General information

Dried or affixed surgical residues can make cleaning more difficult or inef-
fective and lead to corrosion. Therefore the time interval between appli-
cation and processing should not exceed 6 h; also, neither fixating pre-
cleaning temperatures >45 °C nor fixating disinfecting agents (active
ingredient: aldehydes/alcohols) should be used.
Excessive measures of neutralizing agents or basic cleaners may result in
a chemical attack and/or to fading and the laser marking becoming
unreadable visually or by machine for stainless steel.
Residues containing chlorine or chlorides e.g. in surgical residues, medi-
cines, saline solutions and in the service water used for cleaning, disinfec-
tion and sterilization will cause corrosion damage (pitting, stress corro-
sion) and result in the destruction of stainless steel products. These must
be removed by rinsing thoroughly with demineralized water and then dry-
ing.
Additional drying, if necessary.
Only process chemicals that have been tested and approved (e.g. VAH or
FDA approval or CE mark) and which are compatible with the product's
materials according to the chemical manufacturers' recommendations
may be used for processing the product. All the chemical manufacturer's
application specifications must be strictly observed. Failure to do so can
result in the following problems:
Optical changes of materials, e.g. fading or discoloration of titanium or
aluminum. For aluminum, the application/process solution only needs
to be of pH >8 to cause visible surface changes.
Material damage such as corrosion, cracks, fracturing, premature aging
or swelling.
Do not use metal cleaning brushes or other abrasives that would dam-
age the product surfaces and could cause corrosion.
Further detailed advice on hygienically safe and material-/value-pre-
serving reprocessing can be found at www.a-k-i.org, link to Publica-
tions, Red Brochure – Proper maintenance of instruments.
7.4
Disassembling the product before carrying out the
reprocessing procedure
Disassemble the product immediately after use, as described in the
respective instructions for use.
Open up products with hinges.
14
®
7.5

Preparations at the place of use

If applicable, rinse surfaces that are not accessible to visible inspection
(preferably with demineralized water), using a disposable syringe, for
example.
Remove any visible surgical residues as much as possible with a damp,
lint-free cloth.
Place the dry product in a sealed container and forward it on for clean-
ing and disinfection within 6 hours.
7.6

Preparation before cleaning

Disassemble the product prior to cleaning, see Disassembling.
7.7

Cleaning/disinfection

7.7.1

Product-specific safety notes on the reprocessing procedure

Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
CAUTION
Use cleaning and disinfecting agents according
to the manufacturer's instructions which
– are approved for plastic material and stain-
less steel,
– do not effect plasticizers (e.g. in silicone).
Observe specifications regarding concentration,
temperature and exposure time.
Do not exceed the maximum allowable cleaning
temperature of 60 °C.
Carry out ultrasound cleaning:
– as an effective mechanical supplement to manual cleaning/disin-
fecting.
– as a pre-cleaning procedure for products with encrusted residues,
in preparation for mechanical cleaning/disinfecting.
– as an integrated mechanical support measure for mechanical
cleaning/disinfecting.
– for additional cleaning of products with residues left after mechan-
ical cleaning/disinfecting.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis