6 Running the Cables

6
Running the Cables
J
5.
Running the monitor cables
i
NOTE
In the procedure described below for running the cables, the
disinfection tank must only be disassembled for mounting
positions 2 and 3 (cf. diagram on 5). The disinfection tank
does not need to be disassembled for mounting positions 1
and 4. In this case, the cables are run in front of the tank.
ATTENTION
Cables may break! Do not sharply bend or kink XGA cable
L35 or S-video cable L7 !
For mounting positions 2 and 3 only
1.
Remove the screw J and take out the cover and the re-
filling funnel.
2.
Remove the disinfection tank (for DENTOSEPT):
- Loosen the rubber hose on the tank
- Set the disinfection tank down
- Separate the suction hose from suction valve MV33.
12
For positions 2 and 3
6.
3.
1.
For positions 1 and 4
4.
2.
6.
3.
Place the cables in the water unit and run them along
the side of the rinsing tank through the cable tie.
Only for mounting positions 1 and 4 only:
4.
Run the cables on the left side of the disinfection tank to
the connection flange of the water unit.
For all mounting positions:
5.
Run cables L7, L35 and L40 and the protective ground
wire through the cable tie up to the connection box.
6.
Reinstall the disinfectant tank (for installation in Pos. 2
or 3) and suction hose.
Sirona Dental Systems GmbH
59 14 200 D 3322