Starting up

8
Starting up
42
Comply with safety instructions during installation, start­
ing up and operation.
WARNING
Any contamination of the charge amplifier inputs and out­
puts can cause drift, a reduction in insulation resistance,
or even short circuits, which can invalidate the measure­
ment results. This can cause danger for the operating
personnel of the system in which the transducer is in­
stalled.
Protect the inputs and outputs of the charge amplifier
from contamination and do not touch the contacts in the
plug with your fingers.
Recommended cleaning agents: a cleaning pad (e.g.
HBM ordering number 1-8402.0026), pure isopropanol
(e.g. IPA200 from RS Components). Also refer to Section
14.3 on page 72.
How to proceed.
1. Connect the sensors (BNC connection, SENSOR);
charge cable 1-KAB143-3 plus adapter 1-CON-
P3001.
2. Connect the supply voltage (18  30 V
socket);
connection cable 1-KAB168-5(20).
3. Connect the Ethernet interface to a PC (see Sec­
tion 8.2.3);
Ethernet cable 1-KAB284-2.
A2781-3.1 HBM: public
, SYSTEM
DC
CMD