Device description
Parameter
Transducer elec­
trical signal type
Minimum force/
weight
Maximum force/
weight
Minimum electrical
value
Maximum elec­
trical value
Mapping method
AC or DC coupling
Output impedance
of the sensor
Response time
Excitation level
(nominal)
Excitation level
(minimum)
Excitation voltage
type
Max. current draw
at nominal excita­
tion level
38
manufactured on 27.6.2011 at HBM. If applicable, you
can create the required High Level Voltage template.
Template: High Level Voltage
Value
Unit
Sensor
voltage
0.000
N
20.000 k
N
0.00000
V/V
+9.5700
V/V
Linear
DC
10.0
ohm
1.0000000
sec
u
24.0
V
18.0
V
DC
50.12 m
A
A2781-3.1 HBM: public
Authoriz‐
ation
level
ID
CAL
The physical measurand and
unit are defined when the tem­
plate is created, after which
CAL
they cannot be changed.
CAL
The difference between these
values is the sensitivity accord­
ing to the HBM manufacturing
CAL
certificate, or from calibration.
This entry cannot be changed.
ID
ID
Output resistance according to
the CMD data sheet.
ID
Not relevant.
ID
Supply voltage according to the
CMD data sheet.
ID
Lower limit of the supply voltage
operating range according to
the CMD data sheet.
ID
Type of supply voltage
ID
Maximum supply current
Explanation