6

Starting up

Comply with safety instructions during installation, starting up and operation.
CAUTION
Contaminatioon of the charge amplifier inputs and outputs can cause
drifts, reduced insulation resistance and thus short circuits and
incorrect test results.
Protect the inputs and outputs of the charge amplifier from contamination and
do not touch the socket contacts with the fingers.
Recommended cleaning agents: cleaning pad, pure isopropanol (e.g. IPA200
from RS Components). See also chapter 11.3
How to proceed.
1.
Connect the sensors (BNC connection, SENSOR);
charge cable 1−KAB143−3 plus adapter 1−CON−P3001
2.
Connect the supply voltage (18−30 VDC, socket SYSTEM)
connection cable 1−KAB168−5(20)
3.
Connect the Ethernet system interface to a PC (see Chapter 6.2.3 );
Ethernet cable 1−KAB284−2
4.
Connect the digital input/output with e.g. the PLC (see Chapter 6.2.4 )
connecting socket 1−CON−S1002 (optional)
5.
Install the parameter software (see Chapter 6.2.5 )

6.1 CMD600 power up behavior

On power up of the CMD600, the output voltage is −10V and both digital
outputs are on logical state 0 − inactive.
Power up time is 300 ms. After this time all outputs are stable. Power up time
is extended to 13 s, if Sensorteach input is active on power up (see section for
firmware upgrade).