Piezoelectric
transducer
CMD
7.6
Principle of operation
CMD charge amplifier
Fig. 7.1
Block diagram for the CMD measuring chain
The CMD charge amplifier is connected to a transducer
for signal conditioning. The CMD amplifies the electric
charges to a proportional output signal of ‐10 ... +10 V
(see the manufacturing certificate for accurate data).
The RESET/MEASURE input can be used to trigger
measurement, or to clear (RESET) the charge amplifier
input capacitors.
The charge amplifier is fitted with TEDS (Transducer
Electronic Data Sheet).
A2781-3.1 HBM: public
Device description
Supply voltage 18 ... 30 V
SensorTeach
RESET / MEASURE
Digital input 1
Digital output 1
Digital output 2
Ethernet interface
Signal output
-10 ... +10 V
0 V
DC
TEDS
DC
DC
DC
25