Calibration date
Calibration initials
Calibration period
(days)
Measurement loc­
ation ID
CMD
27‐Jun‐20
11
HBM
0
days
0
CAUTION
When the sensor is replaced or the system amplification
changes, the downstream systems will work with the
wrong sensor settings if the TEDS data is not adjusted.
This can cause danger for the operating personnel of the
system in which the transducer is installed.
So the TEDS data must be adjusted when a sensor is
replaced or the amplification changes.
A2781-3.1 HBM: public
CAL
Date of the last calibration or
creation of the manufacturing
certificate (if no calibration car­
ried out), or of the storage of the
TEDS data (if only nominal
(rated) values from the data
sheet were used). Format:
day‐month‐year.
Abbreviations for the months:
Jan, Feb, Mar, Apr, May, Jun,
Jul, Aug, Sep, Oct, Nov, Dec.
CAL
Initials of the calibrator or cali­
bration laboratory concerned.
CAL
Time before recalibration, calcu­
lated from the date specified
under Calibration Date.
USR
Identification number for the
measuring point. Can be
assigned according to the appli­
cation. Possible values: 0 to
2047. If this is not enough for
you, you can also use the HBM
"Channel Comment" template
for this purpose.