6.9
Inspection, maintenance and checks
Allow the product to cool down to room temperature.
Inspect the product after each cleaning and disinfecting cycle to be sure it is: clean, functional, and undamaged.
micro-Line Prior to sterilization, spray through the handpiece with Aesculap-STERILIT oil spray GB600 with
adapter GB600810 for approx. 1 s.
Check the product for any damage, abnormal running noise, overheating or excessive vibration.
Inspect tools for broken, damaged or blunt edges.
Set aside the product if it is damaged.
Prior to an extended period of non-use, store attachments cleaned, maintained and dried according to the
instructions.
6.10 Packaging
Follow the instructions for use for the applied packaging and storage systems (e.g. instructions for use TA009721
for Aesculap Eccos storage system).
Insert the product in its proper position in the Eccos holder, or put it on a tray in such a way that the product is
protected against damage. Ensure that all cutting edges are protected.
Pack trays appropriately for the sterilization process (e.g. in Aesculap sterile containers).
Ensure that the packaging will prevent a recontamination of the product.
6.11 Steam sterilization
Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
opening any valves and faucets).
Validated sterilization process
– Steam sterilization using fractionated vacuum process
– Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665
– Sterilization using fractionated vacuum process at 134 °C/holding time 5 min
When sterilizing several products at the same time in a steam sterilizer, ensure that the maximum load capacity
of the steam sterilizer specified by the manufacturer is not exceeded.
6.12 Sterilization for the US market
Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization.
Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave.
-6
To achieve a sterility assurance level of 10
, Aesculap recommends the following parameters:
Aesculap Orga Tray/Sterile container (perforated bottom)
Minimum cycle parameters*
Sterilization method
Temp.
Prevacuum
270 °F/275 °F
*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an
Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles
may also be suitable, however individuals or hospitals not using the recommended method are advised to validate
any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility
after processing, such as a wrap, pouch, etc.
6.13 Storage
Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
7.

Maintenance

To ensure reliable operation, the product must be maintained at least once a year.
For technical service, please contact your national B. Braun/Aesculap agency, see Technical Service.
8.

Troubleshooting list

Malfunction
Detection
micro-Line straight
Collet not locked
handpiece/angled hand-
piece not running
Marks on the tension ring
and on the handle part no
aligned with each other
Drive unit/motor not
reacting
Insufficient power
Tool not cutting correctly
Drive unit/motor not run-
ning correctly
Tool change impossible
Tension ring does not
rotate freely or does not
rotate at all
Tool blocked
micro-Line straight
Irregular running noise
handpiece/angled hand-
piece vibrating
Tool shaft bent
micro-Line straight
Rapid temperature rise
handpiece/angled hand-
piece heating up during
operation
Time
Minimum drying time
4 min
20 min
Cause
Remedy
Tension ring fixed/closed
Turn the tension ring
or not tightened
against the direction of
the arrow until the marks
on the tension ring and
on the handle part are
aligned with each other
Tension ring fixed/closed
Turn the tension ring
or not tightened
against the direction of
the arrow until the marks
on the tension ring and
on the handle part are
aligned with each other
Drive unit/motor defec-
Check drive unit/motor
tive
according to appropriate
instructions for use
Cutting edges worn
Replace the tool
Drive unit/motor defec-
Check drive unit/motor
tive
according to appropriate
instructions for use
Collet jammed
Have repaired by manu-
facturer
Tool is jammed
Have repaired by manu-
facturer
Bearing defective
Have repaired by manu-
facturer
Tool damaged
Replace the tool
Bearing defective
Have repaired by manu-
facturer
9.
Technical Service
Risk of injury and/or malfunction!
Do not modify the product.
WARNING
For service and repairs, please contact your national B. Braun/Aesculapagency.
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and for-
feiture of applicable licenses.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Or in the US:
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone:
+1 (800) 214 -3392
Fax:
+1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
10.
Accessories/Spare parts
Art. no.
Designation
GD460R
Spray nozzle for GD450M, GD455M, GD465M
GD461R
Spray nozzle for GD451M, GD456M, GD466M
GD462R
Spray nozzle for GD457M, GD467M
GB600
Aesculap STERILIT oil spray
GB600810
STERILIT micro-Line adapter
GB681R
Eccos holder (for 3 handpieces)
GD682R
Eccos holder (for 1 handpiece)
GB459R
Eccos mounting set for holders
TA012581
Instructions for use for GD450M, GD451M, GD455M, GD456M,
GD457M, GD465M, GD466M, GD467M
see Aesculap catalog Power Systems -
Tools (e.g. GD025R)
Burrs & Blades
11.

Technical data

Classification acc. to Directive 93/42/EEC
Art. no.
Designation
GD450M, GD451M,
micro-Line straight handpiece/angled handpiece
GD455M, GD456M,
GD457M, GD465M,
GD466M, GD467M
micro-Line straight hand-
Gear
piece/angled handpiece
ratio
GD450M
1:1
GD451M
1:1
GD455M
1:2
GD456M
1:2
GD457M
1:2
GD465M
1:1
GD466M
1:1
GD467M
1:1
Class
IIa
Motor
Tool speed
Weight
Tool
drive rpm
(g)
length
(rpm)
(mm,
(rpm)
max.)
0–40 000
0–40 000
69
70
0–40 000
0–40 000
76
95
0–40 000
0-80 000
90
70
0–40 000
0-80 000
97
95
0–40 000
0-80 000
104
125
0–40 000
0–40 000
90
70
0–40 000
0–40 000
97
95
0–40 000
0–40 000
104
125
Class
II a
II a
II a
II a
II a
II a
II a
II a