8.2.2 Error Indication: Attendant Control

LED indicator
Red is flashing
Battery under voltage
Green is flashing
Battery over voltage
Mode LED (see Fig. 40, item 4)
Red is flashing 1 x
Controller fault
Hand control device
Red is flashing 2 x
fault
Red is flashing 3 x
Communication error
Red is flashing 4 x
Joystick warning
Brake fault
Red is flashing 5 x
Drive motor fault
C1000 DS
Warning/Error
Battery deep discharge
Voltage too high (after full charge and
driving downhill)
Defective controller
Defective joystick
Faulty connection between the hand
control device and controller;
Defective cabling, software or hardware
Joystick is not in zero position when the
unit is turned on
Open brake release;
Defective brake
Faulty cabling or plug connection;
defective actuator
Cause
Charge as soon as possible
Continue driving slowly
Contact specialist dealer
Contact specialist dealer
Check cabling / plug connec-
tions;
Contact specialist dealer
Move the joystick to the zero
position before turning the unit on
Close brake release;
Check brake (e.g. Bowden cable)
Check cabling / plug connections;
Contact specialist dealer
Malfunctions / troubleshooting