10.3.5.2 Configure digital input (sNDI1)
The digital output DO1 of the LLB measuring system can also be used as digital input. The command
sNDI1 configures an action for the measuring system. The state of the digital input can be read with
the command sNRI
On active digital input, the digital output function of DO1 is deactivated.
Command
Return successful
Return Error
Parameters
Configuration parameters must be saved (see 10.3.8 "Save configuration parameters (sNs)" on page 145).
1) Uses time for the tracking as set by a previous sNh+ or sNuh+ command
10.3.5.3 Read digital input (sNRI)
Displays the state of the digital input, if the input is not inactive.
Command
Return successful
Return Error
Parameters
Printed in the Federal Republic of Germany
03/09/2016
Set Command
sNDI1+xxxxxxxx<trm>
gNDI1?
<trm>
gN@Ezzz<trm>
Device ID (0..9)
N
0: Digital input inactive (DO1 active)
xxxxxxxx
1: activate the digital input to read out its state with the
command 10.3.5.3 "Read digital input (sNRI)".
2: Trigger Distance measurement (sNg)
3: Start/Stop Single sensor Tracking (sNh)
4: Start/Stop Tracking with buffering (sNf)
5: Trigger User Distance measurement (sNug)
6: Start/Stop Single sensor User Tracking (sNuh)
7: Start/Stop User Tracking with buffering (sNuf)
8: Start/Stop Single sensor Timed Tracking (sNh+)
9: Start/Stop Single sensor User Timed Tracking (sNuh+)
Error code
zzz
Command
sNRI<trm>
gNRI+x
<trm>
gN@Ezzz<trm>
Device ID (0..9)
N
0: Input Off (Signal Low)
x
1: Input On (Signal High)
Error code
zzz
TR - ELE - BA - DGB - 0021 - 06
Get Command
sNDI1<trm>
sNDI1+xxxxxxxx<trm>
gN@Ezzz<trm>
TR-Electronic GmbH 2011, All Rights Reserved
1)
1)
Page 141 of 156