10.3.4.2 Set/Get analog output value in error case (sNve)

This command sets the analog output current level in mA in case of an error. This level can be lower
than the minimum level set in 10.3.4.1 "Set/Get analog output min level (sNvm)" on page 138.
Command
Return successful
Return Error
Parameters
Configuration parameters must be saved (see 10.3.8 "Save configuration parameters (sNs)" on page 145).

10.3.4.3 Set/Get distance range (sNv)

Put the minimum and maximum distance into dependence of the minimal and maximum output current
value of the analog output.
0...20 mA
DIST - D
Aout =
D
- D
max
min
Aout:
Analog current output
DIST:
Actual measured distance
D
:
Distance programmed for the minimum output current
min
D
:
Distance programmed for the maximum output current
max
Command
Return successful
Return Error
Parameters
Configuration parameters must be saved (see 10.3.8 "Save configuration parameters (sNs)" on page 145).
Printed in the Federal Republic of Germany
03/09/2016
Set Command
sNve+xxx<trm>
gNve?
<trm>
gN@Ezzz<trm>
Device ID (0..9)
N
Value in case of an error in 0.1 mA
xxx
If the value is set to 999, in case of an error keep the last valid distance.
Error code
zzz
min
* 20 mA
Set Command
sNv+xxxxxxxx+yyyyyyyy<trm> sNv<trm>
gNv?
<trm>
gN@Ezzz<trm>
Device ID (0..9)
N
Minimum distance in 1/10 mm corresponding to 0 mA / 4 mA
xxxxxxxx
Maximum distance in 1/10 mm corresponding to 20 mA
yyyyyyyy
Error code
zzz
TR - ELE - BA - DGB - 0021 - 06
Get Command
sNve<trm>
sNve+xxx<
<trm>
gN@Ezzz<trm>
4...20 mA
DIST - D
min
Aout =
D
- D
max
min
Get Command
gNv+xxxxxxxx+yyyyyyyy
gN@Ezzz<trm>
 TR-Electronic GmbH 2011, All Rights Reserved
* 20 mA + 4 mA
<trm>
Page 139 of 156